SONARtrac® PROCESS MONITORING SYSTEM QUICK START GUIDE CiDRA Corporation Tel: 203 265 003550 Barnes
Menu Mode Key Operational Mode Navigation Editing Dialog Box Up Enter Menu Mode Cycle Menu Item Change current value at cursor position Exit Dialog
• Press the ‘ENTER’ key to start the sensor test. Each of the sensors will be tested. If any sensor does not PASS, repeat the test. Individual se
3.2.1.2.2 Check / Set Gain Check / Set Gain allows the user to check the current gain setting in the pre-amplifier and to manually set it to one of
• Press the ‘ENTER’ key to access the options available under that menu. • ‘Autoset Gain’ will appear on Line 3 of the display. • Press the ↓ key
3.2.2 Process Off-Line Some times the SONARtrac system is installed with the process off-line. In this case, it is not possible to verify the prope
3.3 Operating Menu Setup 3.3.1 VF-100 System Initial Setup In order to conduct measurements, the ‘Basic Config’ setup must be completed. If power
• Following the entering of ‘Pipe Size’ press the ‘BACK’ key and ‘→Pipe Size’ will be displayed on Line 2. At this point the user can either press
desired format. Press the ‘ENTER’ key to set the desired date format followed by the ‘BACK’ key. • At this point ‘Basic Config’ is displayed on Lin
• The third choice on the ‘Pipe Size’ menu is ‘OD / Wall’ displayed on Line 3. Use the arrow keys (↑ and ↓ to scroll through the values and ← and →
arrow keys to enter the new value. This will result in a fixed pressure being used for GVF calculations. If a Pressure Transducer will be used to i
• Pressing the ↓ key will next display ‘→Calibration’ on Line 2. Press the ‘ENTER’ key and ‘→C0’ will appear on Line 3 and a numerical value on Line
4 HART COMMUNICATIONS 4.1 Introduction The SONARtrac transmitter implementation of HART supports HART Universal Command Revisions 5 and 6 as well as
4.2.2 4-20mA Transmitter and DCS Connections Connect to the appropriate terminals on the transmitter connector board. Figure 4 Transmitter Co
4.2.4 HART Activity Indicator When the SONARtrac receives a valid HART message, an indicator will appear on the lower right of the SONARtrac display
4.2.5.3 Preambles/Response Preambles The number of Preambles or Response Preambles should not need to be changed unless the 4-20mA current loop is no
4.2.6.9 PV Trim Value subtracted from the HART ‘Primary Variable’ (i.e. VF). 4.3 HART Troubleshooting 4.3.1 HART Master Does Not Connect to SONARtra
4.4 HART Commands The following is a list of supported HART commands. Command Name HART 5 or 6 Supported by CiDRA Universal Commands 0 Read uniq
50 Read dynamic variable assignments (4.1) HART 5 & 6 NO 51 Write dynamic variable assignments (4.1) HART 5 & 6 NO 52 Set transmitter vari
*** This Page is Blank ***Copyright © 2007 CiDRA Corporation Page 4-820984-01 Rev 01
Table of Contents 1 INTRODUCTION...1-1
CiDRA Corporation 50 Barnes Park North Wallingford, CT 06492 CiDRA Corporation 50 Barnes Park North Wallingford, CT 06492 (In U.S.):
*** This Page is Blank *** Copyright © 2007 CiDRA Corporation Page ii20984-01 Rev 01
1 INTRODUCTION 1.1 Purpose The purpose of this document is to provide an overview of the CiDRA SONARtrac system startup and operation. It is not in
*** This Page is Blank ***Copyright © 2007 CiDRA Corporation Page 1-220984-01 Rev 01
2 TRANSMITTER LAYOUT 2.1 Transmitter Illustration The following figure illustrates the SONARtrac transmitter. Navigation ke
*** This Page is Blank ***Copyright © 2007 CiDRA Corporation Page 2-220984-01 Rev 01
3 TRANSMITTER STARTUP AND OPERATION 3.1 Initial Start-up At initial power up the Green LED will light and the transmitter will display startup progr